82
STOREF_n?
Store Reference Reading Query
Syntax:
STOREF_n?
where n is either A or B
Parameters:
None
Function:
This query returns the latest reading stored as a reference reading. The units
of the reference reading defaults to Watts for Low-Power and High-Power
detectors and Joules for Energy detectors.
Returns:
<number>
<number> is a floating point number in the exponential format.
Related Commands:
REFSEL_n, REFSEL_n?, USRREF_n, USRREF_n?, STOREF_n
STOZERO_n
Store Zero Signal
Syntax:
STOZERO_n
where n is either A or B
Parameters:
None
Function:
This command takes the latest reading and stores it as a zero signal reference
value to be used when zeroing is enabled. The units of the zero reference
defaults to Amperes for Low-Power detectors and Volts for High-Power and
Energy detectors.
Related Commands:
ZERO_n, ZERO_n?, ZEROVAL_n?
TONE
Tone Enable
Syntax:
TONE <state>
Parameters:
The parameter <state> is <number> which can be 0 or 1. If the number is 0
then the tone is disabled. If the number is 1, then tone is enabled.
Function:
This command enables or disables a tone feature that emits a short audible
“tick” when a peak or pulse is detected in peak-to-peak or pulse acquisition
modes. This command will cause an error if the detector does not support
peak-to-peak or pulse acquisition modes.
Related Commands:
TONE?