6
the channel indicated by the display channel annunciators. Table 2 and
Sections 2.3.1 through 2.3.21 list and describe each top level key function.
Menu level key functions are discussed in Section 2.4.
R/S B
CH B
R/S A
CH A
STO REF
DISP
RANGE
AUTO
ENTER
UNITS
REF SEL
FILTER
MENU
MODE
O I
(LOCAL)
SHIFT
ESC
STATS
EXT
R/S
λ
ZERO
Figure 3 – Front Panel Key Pad
Table 2 – Top Level Key Functions and Associated Remote Commands
Keypad
Remote Commands
Description
O I
None
Turns the Model 2835-C on and off.
SHIFT
None
Enables the blue key functions.
DISP
DISP, DISP?
Cycles display brightness:
OFF,LOW,NORM and HIGH.
FILTER
FILTER_n, FILTER_n?
Cycles signal filtering:
OFF, ANLG, DIG and ANLG+DIG.
ZERO
ZERO_n, ZERO_n?,
Zeroes the display via offset subtraction.
STOZERO_n, ZEROVAL_n?
AUTO
AUTO_n, AUTO_n?
Turns automatic gain ranging on and off.
CH A
DISPCH, DISPCH?
Selects channel A as the display channel.
CH B
DISPCH, DISPCH?
Selects channel B as the display channel.
STO REF
STOREF_n, STOREF_n?
Stores last reading for future dB and
REL measurements.
REF SEL
REFSEL_n, REFSEL_n?
Selects the reference source to be used
for dB and REL measurements.
λ
LAMBDA_n, LAMBDA_n?
Displays and edits the calibration
wavelength in use.
RANGE
RANGE_n, RANGE_n?
Displays signal gain range in use.
,
None
Adjusts gain range or parameter
values or moves through a list.
R/S
RUN, STOP
Toggles the signal acquisition of
both channels.
MODE
MODE_n, MODE_n?
Cycles meter between allowed
measurement modes.
UNITS
UNITS_n, UNITS_n?
Cycles display measurement units
between those allowed.
STATS
STSIZE_n, STSIZE_n?, STMAX_n?
Displays statistics.: Max, Min,
STMIN_n?, STMXMN_n?
Max-Min, Mean and Std. Dev.
STMEAN_n, STSDEV_n?
R/S A
RUN_A, RUN_A?
Starts and stops channel A signal
acquisition.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com