81
STSIZE_n?
Statistics Buffer Size Query
Syntax:
STSIZE_n?
where n is either A or B
Parameters:
None
Function:
This query returns the statistics buffer size.
Returns:
<size>
<size> is of type <number> and represents an integer in the range 0 to 100.
Related Commands:
STMAX_n?, STMEAN_n?, STMIN_n, STMXMN_n?,
STSDEV_n?, STSIZE_n
STOP
Stop Both Channel Acquisition
Syntax:
STOP
Parameters:
None
Function:
This command terminates any acquisition of data currently in progress on
both channels. If data is not being acquired on a channel then this command
has no effect.
Related Commands:
RUN, RUN_n, STOP_n
STOP_n
Stop Single Channel Acquisition
Syntax:
STOP_n
where n is either A or B
Parameters:
None
Function:
This command terminates any acquisition of data currently in progress on the
specified channel. If data is not being acquired on that channel the command
has no effect.
Related Commands:
RUN, RUN_n, STOP
STOREF_n
Store Reference Reading
Syntax:
STOREF_n
where n is either A or B
Parameters:
None
Function:
This command takes the latest reading and stores it as a reference reading to
be used when making relative linear and dB measurements. The units of the
reference reading defaults to Watts for Low-Power and High-Power detectors
and Joules for Energy detectors.
Related Commands:
REFSEL_n, REFSEL_n?, USRREF_n, USRREF_n?, STOREF_n?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com