71
LAMBDA_n
Lambda Select
Syntax:
LAMBDA_n <wavelength>
where n is either A or B
Parameters:
The parameter <wavelength> is of type <number> which rounds to an integer
with a range that depends on the detector being used. The units of <wave-
length> are nanometers and should correspond to the wavelength of the light
source being measured.
Function:
This command is used to specify the wavelength of light being measured.
This will insure that the proper responsivity calibration point will be used
when the 2835-C calculates measurement values.
Related Commands:
ATTN_n, ATTN_n?, LAMBDA_n?, RESP_n?, USRCAL_n,
USRCAL_n?, USRRESP_n, USRRESP_n?
LAMBDA_n?
Lambda Query
Syntax:
LAMBDA_n?
where n is either A or B
Parameters:
None
Function:
This query returns the wavelength that corresponds to the responsivity calibra-
tion point currently being used by the 2835-C in measurement calculation.
Returns:
<wavelength>
<wavelength> is of type <number> and represents an integer. The units of
<wavelength> are nanometers.
Related Commands:
ATTN_n, ATTN_n?, LAMBDA_n, RESP_n?, USRCAL_n,
USRCAL_n?, USRRESP_n, USRRESP_n?
LLO
Local Lockout Enable
Syntax:
LLO <state>
Parameters:
The parameter <state> is either 0 or 1. If <state> is 0 then local lock-out is
disabled. If <state> is 1 then local lock-out is enabled.
Function:
This command enables or disables the local lockout.
Returns:
None
Related Commands:
LLO?