86
TRIGOUTPOL
Trigger Out Polarity Select
Syntax:
TRIGOUTPOL <state>
Parameters:
The parameter <state> is a <number> which can be 0 or 1. If the number
rounds to 0, then the polarity will be active low. If the number is 1, then the
polarity will be active high.
Function:
This command is used to set the polarity of the trigger out output to either
active low or active high.
Related Commands:
TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTINTG, TRIGOUTINTG?, TRIGOUTLVL, TRIGOUTLVL?, TRIGOUTPOL?
TRIGOUTPOL?
Trigger Out Polarity Query
Syntax:
TRIGOUTPOL?
Parameters:
None
Function:
This query returns a value that indicates whether the trigger out polarity is
active high or active low.
Returns:
<state>
<state> is of type <number> and represents the integer 0, if the polarity is
active low, or 1, if the polarity is active high.
Related Commands:
TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTINTG, TRIGOUTINTG?, TRIGOUTLVL, TRIGOUTLVL?, TRIGOUTPOL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com