Performance Considerations
88
O
XI
Max
N-65 Operator’s Manual
Dysfunctional Hemoglobins
Dysfunctional hemoglobins, such as,
carboxyhemoglobin, methemoglobin, and
sulfhemoglobin, are unable to carry oxygen. SpO
2
readings may appear normal; however, a patient may be
hypoxic because less hemoglobin is available to carry
oxygen. Further assessment beyond pulse oximetry is
recommended.
Anemia
Anemia causes decreased arterial oxygen content.
Although SpO
2
readings may appear normal, an anemic
patient may be hypoxic. Correcting anemia can improve
arterial oxygen content. The monitor may fail to provide
an SpO
2
if hemoglobin levels fall below 5 gm/dl.
Saturation
The monitor displays saturation levels between 1% and
100%.
Pulse Rates
The monitor displays pulse rates between 20 and 300
beats per minute. The sensor accuracy ranges do not
apply to pulse rates above 250 bpm. Detected pulse rates
below 20 are shown as 0.