failure to cover the
O
XI
M
AX
sensor site with opaque
material in high ambient light conditions
•
venous pulsation
•
dysfunctional hemoglobin
•
low perfusion
Loss-of-pulse signal can occur for the following reasons:
•
the
O
XI
M
AX
sensor is applied too tightly
•
defibrillation
•
a blood pressure cuff is inflated on the same
extremity as the one with the
O
XI
M
AX
sensor
attached
•
there is arterial occlusion proximal to the
O
XI
M
AX
sensor
•
poor peripheral perfusion
•
loss of pulse/cardiac arrest
High ambient light sources that can interfere with the
performance of an SpO
2
O
XI
M
AX
sensor are:
•
surgical lights (especially those with a xenon light
source)
•
bilirubin lamps
•
fluorescent lights
•
infrared heating lamps
•
direct sunlight