Caution: The
O
XI
M
AX
Sensor Disconnected icon and
associated alarm indicate that either the
O
XI
M
AX
sensor is disconnected or the wiring is faulty. Check
the
O
XI
M
AX
sensor connection and, if necessary,
replace the
O
XI
M
AX
sensor, extension cable, or both.
For additional information on monitor measurements, see
Inaccurate Monitor Measurements
on page 28.
For a complete current list of all
O
XI
M
AX
sensors
applicable to the
O
XI
M
AX
N-65 refer to the latest Sensor
Accuracy Grid posted on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
When selecting an
O
XI
M
AX
sensor consider the following:
•
patient weight and activity
•
adequacy of perfusion
•
available
O
XI
M
AX
sensor sites
•
need for sterility
•
anticipated duration of monitoring
Содержание OxiMax N-65
Страница 1: ......
Страница 12: ...Safety Information 6 OXIMax N 65 Operator s Manual ...
Страница 14: ...Introduction 8 OXIMax N 65 Operator s Manual ...
Страница 36: ...Set up the OXIMAX N 65 30 OXIMax N 65 Operator s Manual ...
Страница 92: ...OXIMAX Sensors and Accessories 86 OXIMax N 65 Operator s Manual ...
Страница 98: ...Performance Considerations 92 OXIMax N 65 Operator s Manual ...
Страница 108: ...Troubleshooting 102 OXIMax N 65 Operator s Manual ...
Страница 112: ...Maintenance 106 OXIMAX N 65 Operator s Manual ...
Страница 118: ...Menu Structure 112 OXIMAX N 65 Operator s Manual ...
Страница 150: ......
Страница 151: ......