Turn On the
O
XI
M
AX
N-65
Discussion
Verify the monitor works properly and is safe to use.
Proper operation of the monitor is verified each time it is
turned on as described in this procedure. The verification
procedure Power-On-Self-Test (POST) takes
approximately 10 seconds to complete. When the
monitor is turned on, POST automatically tests the
monitor circuitry and functions.
Caution: During the Power-On-Self-Test (POST)
immediately after power
-
up, confirm that all display
segments and icons are shown and the monitor
speaker sounds a one
-
second tone.
WARNING: If you do not hear the POST pass tone,
do not use the
O
XI
M
AX
N-65.
Alarm Silence
Duration
Alarms 30, 60,
90, 120 seconds,
OFF
30 seconds
Alarm Volume
1 to 10
10
Pulse Beep
Volume
0 to 10
10
Table 2: Parameter Ranges and Defaults (Continued)
Parameter
Ranges/
Selections
Factory
Defaults
Содержание OxiMax N-65
Страница 1: ......
Страница 12: ...Safety Information 6 OXIMax N 65 Operator s Manual ...
Страница 14: ...Introduction 8 OXIMax N 65 Operator s Manual ...
Страница 36: ...Set up the OXIMAX N 65 30 OXIMax N 65 Operator s Manual ...
Страница 92: ...OXIMAX Sensors and Accessories 86 OXIMax N 65 Operator s Manual ...
Страница 98: ...Performance Considerations 92 OXIMax N 65 Operator s Manual ...
Страница 108: ...Troubleshooting 102 OXIMax N 65 Operator s Manual ...
Страница 112: ...Maintenance 106 OXIMAX N 65 Operator s Manual ...
Страница 118: ...Menu Structure 112 OXIMAX N 65 Operator s Manual ...
Страница 150: ......
Страница 151: ......