Battery Operation
32
O
XI
Max
N-65 Operator’s Manual
Caution: 15 minute indicator for remaining battery
operating time is approximate and is based on
Alkaline AA batteries. Remaining operating time may
be different for other types of batteries.
Caution: Periodically check the battery for corrosion.
If the
O
XI
M
AX
N-65 is to be stored for three months or
longer, remove the batteries from the monitor before
storage.
Critical Battery Indication
When the batteries are critically low, the monitor:
•
displays an error message (Er 521)
•
sounds a high priority alarm
•
shuts down
Replace the batteries and restart the monitor. See
Battery
Installation
on page 33.