Specifications
O
XI
Max
N-65 Operator’s Manual
125
High Priority
Alarm
Volume level
Adjustable, 42 to 52
dB(A), at one meter
Pitch (±30 Hz)
1200 Hz
On pulse width
(±20 msec)
250 msec
Off Interval (±20 msec)
80 msec
Number of pulses in
burst
1
Repetition Pause
(±2 sec.)
0 sec.
Priority
9
Medium Priority
Alarm
Volume level
Adjustable, 42 to 52
dB(A), at one meter
Pitch (±30 Hz)
752 Hz
On pulse width
(±20 msec)
400 msec
Off Interval (±20 msec)
300 msec
Number of pulses in
burst
1
Repetition Pause
(±2 sec.)
0 sec.
Priority
10
Audible Indicators (Continued)
Audible
Indicator
Parameter
Value