Troubleshooting
94
O
XI
Max
N-65 Operator’s Manual
See Table 5, "Error Codes", on page 94 for a list of error
codes and corrective actions.
When an error occurs, the unit:
•
stops monitoring
•
removes all information from the display and
displays the message “Er XXX”
•
sounds a low priority alarm
When an error occurs, you must:
•
Turn the monitor off.
•
Wait 10 seconds and turn the monitor on.
•
If the error appears, follow the action(s) listed in
Table 5.
Table 5: Error Codes
Error
Code
Action
10
Check/replace sensor/extension cable.
11
1 —
Replace batteries.
2 —
Notify service personnel.
17
Check/replace sensor/extension cable.
19
Check/replace sensor/extension cable.
273
1 —
Restart the monitor.
2 — Set the time and date.
3 —
Notify service personnel.
274
Return the monitor for reprogramming.
275
Check/replace sensor/extension cable.
276
Replace with O
XI
M
AX
sensor.