N-180 Pulse Oximeter Service Manual
A-2
Table A-1: Serial Data Output
Character transmitted to
monitor
String received from monitor
R (pulse rate)
Rnnn
S (oxygen saturation)
Snnn
P (not used)
P000
L (low rate alarm limit)
Lnnn
H (high rate alarm limit)
Hnnn
O (low sat. alarm limit)
Onnn
A (alarm status in ASCII
coded decimal
(Logic 1 equals alarm condition)
Bit 0
high rate
Bit 1
low rate
Bit 2
low sat
Bit 3
high sat
M (monitor status)
(In ASCII coded decimal)
Bit Status
1
0
0
Pulse search
Locked
Search
1
Sensor
Attached
Off
2
Audio alarm
Enabled
Disabled
3
ECG
Not in
use
In use
4-7 Not used
—
—
T (time)
T000000
Q (high saturation alarm
limit)
Qnnn