N-180 Pulse Oximeter Service Manual
6-4
Phillips Head
Screw (4)
Simplified bottom view of monitor.
2.
Using a number two Phillips head screwdriver, remove the four bottom
cover assembly screws.
3.
Carefully return monitor to upright position with front panel facing you,
and gather up the screws. Retain for replacement.
Removing Bottom Cover Components
Replaceable components in the bottom cover include:
•
Battery
•
Speaker and speaker brackets
•
Transformer
•
AC inlet (AC power inlet and control)
•
Top shield
•
Oximetry module
•
Middle shield
•
Battery fuse
•
CPU PCB
•
Rear panel
Battery
The following battery removal steps assume you are not troubleshooting the
monitor, in which case, you must leave the battery connected. To remove the
12 VDC battery:
1.
Gain access to the battery by following the instructions in this section for
gaining access to components.
2.
Remove four screws from battery bracket (figure 6-3) and lift off bracket.
3.
Use a small flat blade screwdriver to carefully pry the battery clip
assembly from the battery.
4.
Remove battery.