N-180 Pulse Oximeter Service Manual
7-12
2.
Connect AC plug to electrical safety analyzer as recommended by
operating instructions.
3.
Connect electrical safety analyzer leakage input lead to ground point on
rear panel of monitor.
4.
Verify analyzer indicates 100 microamps or less in on or standby
operating modes for all of the following AC power configurations.
AC polarity
Ground wire
Normal
Normal
Reverse
Normal
Patient Source Current
1.
Set electrical safety analyzer Function to Leakage and Range to
Microamps.
2.
Connect monitor AC plug to electrical safety analyzer as recommended
by analyzer operating instructions for patient source current.
3.
Connect electrical safety analyzer leakage input lead to monitor sensor
input connector, using adapter one.
The analyzer leakage indication must not exceed 10 microamps for 120 V
monitors and 50 microamps for 220 V monitors. These readings apply for
all of the following AC power configurations when the monitor
ON/STDBY switch is set to ON or STDBY.
AC polarity
Ground wire
Normal
Normal
Reverse
Normal
Reverse
Open
Normal
Open
Patient Sink Current
1.
Set electrical safety analyzer Function to Leakage and Range to
Microamps.
2.
Connect monitor AC plug to electrical safety analyzer as recommended
by analyzer operating instructions for patient sink current.
3.
Connect electrical safety analyzer leakage input lead to monitor sensor
input connector, using adapter one.
The analyzer leakage indication must not exceed 10 microamps for 120 V
monitors and 50 microamps for 220 V monitors. These readings apply for
all of the following AC power configurations when the monitor
ON/STDBY switch is set to ON or STDBY.
AC polarity
Ground wire
Normal
Normal
Reverse
Normal
Reverse
Open
Normal
Open