Section 6: Disassembly Guide
6-11
a.
Remove AC power cord.
b.
Insert a screwdriver or other suitable tool into the rear panel slot on
the rear panel (figure 6-8).
c.
Carefully pry the fuse holder door down and out (figure 6-9).
There are two fuse holders behind the fuse holder door.
Fuse drawer (34)
Slot
Equal potential
terminal (86)
Gasket (43)
Dip switch cover
(71)
S/N label
barcode (40)
N-180 Pulse Oximeter
SER. NR.
UL label
(U.S. only)
CSA label
(U.S. only)
Label (87) & (88)
and rear panel (52)
28018A/55
Figure 6-8: AC Power Inlet and Control Assembly
2.
To remove a fuse holder insert the blade of a small screwdriver behind
the white arrow embossed on the fuse holder (not illustrated) and pry
out.
Only one fuse holder is withdrawn in the illustration.
3.
Remove remaining fuse; examine both fuses and determine condition of
each fuse.
4.
Slide a new fuse of the correct rating into the fuse holder that previously
held the blown fuse (follow the legend on the rear panel when
determining correct rating).
5.
Replace remaining fuse.