8-1
SECTION 8: SPECIFICATIONS
Performance
Environmental Requirements
Electrical Characteristics
Physical Characteristics
PERFORMANCE
Range
Saturation
0–100%
Pulse Rate
20–250 bpm (beats per minute)
Accuracy
Saturation
(%SpO
2
±
1 standard deviation)
SpO
2
Adults
70–100 %
±
2 digits
50–69 %
±
3 digits
0–49 %
unspecified
Neonates
70–95 %
±
3 digits
Pulse Rate
20–250 bpm
±
3 bpm
Note:
The monitor pulse oximeter is calibrated to read oxyhe moglobin sat-
uration (%SpO
2
) of functional hemoglobin. Significant levels of
dysfunctional hemoglobins (for example) carboxyhemoglobin, methe-
moglobin) may affect the accuracy of the monitor. Indocyanine green,
methylene blue, and other intra-vascular dyes, depending upon their
con centrations, may interfere with the accuracy of the monitor.
Instrument perfor mance may also be compromised by exces sive pa-
tient movement, electrosurgical interference, or intense environmen-
tal illumination.