Section 5: Troubleshooting
5-15
In table 5-14, the monitor is disconnected from an AC power source and the
ON /STDBY switch has no effect.
Table 5-14: Cannot Turn On Monitor (Battery Operation)
Probable cause
Action
The battery is discharged.
To restart the monitor, set the
ON /STDBY switch to STDBY,
connect the AC power cord to a
known good AC outlet, set the
Power switch to ON , and set the
ON /STDBY switch to ON
The battery is discharged and the
AC power source is defective.
To restart the monitor, set the
ON /STDBY switch to STDBY,
connect the AC power cord to a
known good AC outlet, set the
Power switch to ON , and set the
ON /STDBY switch to ON
Defective monitor
Contact qualified service authority
Defective AC input assembly
Replace the AC input assembly
Defective power transformer
Replace the power transformer
Defective CPU PCB
Replace the CPU PCB
Table 5-15: Cannot Turn Off Monitor
Probable cause
Action
Defective monitor
Contact qualified service authority
Defective ON /STDBY switch
assembly
Replace ON /STDBY switch
assembly
Incomplete connection at CN1 or
CN5 on the LED Drive PCB.
Verify connections at CN1 and CN5
on the LED Drive PCB.
Incomplete connection at CN2 on the
CPU PCB.
Verify connection CN2 on the CPU
PCB.
Defective CPU PCB.
Replace CPU PCB.
Table 5-16: Push Buttons
Probable cause
Action
Defective monitor
Contact qualified service authority
Incomplete connection at CN1 or
CN6 on the LED Drive PCB.
Verify connections at CN1 and CN6
on the LED Drive PCB.
Incomplete connection at CN2 on the
CPU PCB.
Verify connection CN2 on the CPU
PCB.
Defective CPU PCB.
Replace CPU PCB.