N-180 Pulse Oximeter Service Manual
6-12
Fuse drawer (opened)
Fuse holder
Fuse (33)
28018A/60
Figure 6-9: Fuse Holder
Separating Front Panel Assembly From Bottom Cover
The front panel assembly has a groove that fits into the bottom cover housing
on either side. To remove the front panel assembly:
1.
Gain access to the front panel following the instructions for gaining
access to components in this section.
2.
Release ribbon cable retainer clip and disconnect ribbon cable assembly
(figure 6-10) from CN2 of CPU PCB .
3.
Remove front panel ground wire from the chassis by removing the
Phillips head screw located at the front-left battery hold-down bracket.
4.
Release from 10-pin retainer clip holding the sensor cable assembly
connector (to male connector of oximetry module) and disconnect sensor
connector .
6.
Grasp front panel at top edge and lift free of bottom cover.