Section 5: Troubleshooting
5-7
Table 5-5:
Intermittently Lighted
Probable cause
Action
Defective monitor
Contact qualified service authority
Defective connection between the
LED Drive and LED PCBS.
Check the connectors labeled CN on
both PCBs.
Defective leds on the LED PCB.
Replace LED PCB.
Defective display decoder and driver
ICs on the LED Drive PCB.
Replace the LED Drive PCB.
Defective CPU PCB
Replace the CPU PCB
Partial indicator lighting shows up as a dim region behind the panel
membrane.
Table 5-6: Partially Lighted
Probable cause
Action
Defective monitor
Contact qualified service authority
Defective connection between the
LED Drive and LED PCBs.
Check the connectors labeled CN on
both PCBs.
Defective LEDs on the LED PCB.
Replace LED PCB.
Defective display decoder and driver
ICs on the LED Drive PCB.
Replace the LED Drive PCB.
Defective CPU PCB
Replace the CPU PCB
Table 5-7:
Not Lighted
Probable cause
Action
Defective monitor
Contact qualified service authority
Defective connection between the
LED Drive and LED PCBs.
Check the connectors labeled CN on
both PCBs.
Defective LEDs on the LED PCB.
Replace LED PCB.
Defective display decoder and driver
ICs on the LED Drive PCB.
Replace the LED Drive PCB.
Defective CPU PCB
Replace the CPU PCB