Error
Possible Causes / Elimination
The inhalation times are
noticeably longer
and
there is
no or only a small
aerosol output
.
Possible Causes:
-
Dirty or clogged membrane of the SaHoMa
®
-II
mesh nebulization unit
-
Mechanically damaged membrane of the
SaHoMa
®
-II mesh nebulization unit (e. g. in
case of improper handling)
-
The end of the lifespan of the SaHoMa
®
-II
mesh nebulization unit has been reached or
exceeded
-
The room or ambient temperature is below
10° C
Elimination:
-
Cleaning the SaHoMa
®
-II mesh nebulization
unit
(chapter 4.2)
-
Replacement of the SaHoMa
®
-II mesh
nebulization unit
(chapter 9)
-
Please wait approx. 30 minutes until the
system has adjusted to room temperature and
only fill in and nebulize inhalation solutions
with room temperature.
The SaHoMa
®
-II control
unit
switches off
unintentionally
during the
inhalation process despite
a remaining residual
quantity.
Possible Causes:
-
Air bubbles on the membrane of the
SaHoMa
®
-II mesh nebulization unit
-
Use of a foaming medication suggesting a
filling in form of foam residues
Elimination:
-
Light tapping of the SaHoMa
®
-II
mesh nebulization unit to loosen the air
bubbles; if necessary restart the SaHoMa
®
-II
control unit
-
Proper dilution of foaming medications
according to the list of medications;
nebulization in program 2 under constant
observation with manually switching off the
SaHoMa
®
-II control unit.