Cleaning and storing the SaHoMa-II inhalation set for horses
Page
29
Version: NTI_HM-200_IFU-EN_G
Cleaning the SaHoMa
®
-II mesh nebulization unit when using brine inhalants
Especially when brine inhalants are used, the SaHoMa
®
-II mesh nebulization unit may
be clogged by salt crystals.
Therefore, proceed as follows when cleaning:
1.
Fill approx. 5 ml of distilled water into the medication chamber of the
SaHoMa
®
-II mesh nebulization unit.
2.
Set the SaHoMa
®
-II mesh nebulization unit aside for a few minutes and let the
distilled water act on the mesh membrane. During this time it dissolves the salt
particles.
3.
Empty the SaHoMa
®
-II mesh nebulization unit completely and refill approx.
1 ml of distilled water.
4.
Start the SaHoMa
®
-II control unit and wait until the distilled water has been
completely nebulized. Then switch off the SaHoMa
®
-II control unit.
5.
Let the open SaHoMa
®
-II mesh nebulization unit air dry on an absorbent
surface. Make sure that no residual liquid can run back onto the membrane.
☝
Advice
Cleaning with concentrated detergent makes little sense with persistant salt
incrustations. Therefore, these should always be removed as described above
before using the concentrated detergent.