Cleaning and storing the SaHoMa-II inhalation set for horses |
4
Version: NTI_HM-200_IFU-EN_G
Page
30
Cleaning the SaHoMa
®
-II mesh nebulization unit with the concentrated detergent
As soon as the performance of the SaHoMa
®
-II mesh nebulization unit decreases
noticeably, cleaning with the concentrated detergent is recommended. Proceed as
follows:
1.
Fill 2-3 ml of the concentrated detergent into the medication chamber of the
SaHoMa
®
-II mesh nebulization unit.
2.
Start the SaHoMa
®
-II control unit and nebulize a part of the filled
concentrated detergent for about one minute. Then switch off the SaHoMa
®
-II
control unit.
3.
Now wait approx. 3 minutes so that the concentrated detergent can act on
the membrane of the SaHoMa
®
-II mesh nebulization unit.
4.
Then nebulize the remaining concentrated detergent completely as described
in step 2.
5.
Now fill the SaHoMa
®
-II mesh nebulization unit with approx. 3 ml distilled
water from the dosing bottle. Shake the SaHoMa
®
-II mesh nebulization unit
well and then pour out the distilled water.
6.
Finally, nebulize 1 ml of distilled water again.
7.
Let the open SaHoMa
®
-II mesh nebulization unit air dry on an absorbent
surface. Make sure that no residual liquid can run back onto the membrane.
⚠
ATTENTION!
Always ensure optimum ventilation when nebulizing the concentrated detergent!
Although the concentrated detergent is not harmful, it should
not
be inhaled!
⚠
Caution!
Only use the concentrated detergent offered by the manufacturer for cleaning.