Troubleshooting
If you experience problems while using the SaHoMa
®
-II inhalation set for horses, the
following table can help you to find the causes and eliminate them.
Error
Possible Causes / Elimination
The display shows a light
bar, which is on the
left
and the multifunctional
light flashes
alternately
red and green
.
Possible Causes:
-
The SaHoMa
®
-II mesh nebulization unit has
reached the end of its lifespan
-
No SaHoMa
®
-II mesh nebulization unit is
connected
-
There is a cable break in the SaHoMa
®
-II mesh
nebulization unit
-
The battery capacity is too low for inhalation
Elimination:
-
Check the connection of the SaHoMa
®
-II mesh
nebulization unit and the SaHoMa
®
-II control
unit
(chapter 2.8)
-
Replacing the old SaHoMa
®
-II mesh
nebulization unit with a new one
(chapter 9)
-
Charging the SaHoMa
®
-II control unit
(chapter 5.2)
The display shows a light
bar, which is on the
right
and the multifunctional
light flashes
alternately
red and green
.
Possible Causes:
-
Increased power consumption of the
SaHoMa
®
-II mesh nebulization unit
Elimination:
-
To protect the electronic of the SaHoMa
®
-II
control unit, this SaHoMa
®
-II mesh
nebulization unit has to be disposed or used in
program 2 only
(chapter 3.3)
ATTENTION: Further use in program 1 can cause
damage to the SaHoMa
®
-II control and mesh
nebulization unit!