22
23
For a nebulization time of 10 minutes, keep the button pressed until a second bar appears.
In this way, you can set a maximum nebulization time of approx. 25 minutes, which corresponds to five light
bars. If you keep the button pressed, the display starts again from the beginning in the first field and at a time
of five minutes.
When the desired time setting is reached, release the button. The SaHoMa®-II mesh nebulization unit now
produces aerosol as long as you have set it via the SaHoMa®-II control unit. The light bar runs from left to right
and the multifunctional light lights up green constantly.
When the set time has elapsed, the aerosol production stops. The SaHoMa®-II control unit indicates the end of
the inhalation process by two light bars running synchronously from the center to the right and left.
After some time the SaHoMa®-II control unit switches off automatically. The SaHoMa®-II control unit can also be
switched off manually by shortly pressing the button.
Caution
In program 2, there is
no
automatic switch-off when the inhalation solution is completely used up! Therefore,
please make sure that you have filled in an adequate amount of inhalation solution for the selected period
of time or stop the inhalation process manually after the inhalation solution has been completely nebulized.
Running empty can cause damage to the SaHoMa®-II mesh nebulization unit.
Note
The SaHoMa®-II control unit resets itself after being switched off. The desired time must therefore be set again
for the next inhalation process.
3|
Performing and ending the inhalation process
Performing and ending the inhalation process
|3