7.3. Operating conditions
In the following environmental conditions, the proper operation of the components as well as the specified
nebulizer performance and aerosol spectrum are guaranteed:
Technial data
|7
Ambient temperature
+10° C to +30° C; 20° C optionally
Relative air humidity
25 % to 75 %
Air pressure
450 hPa to 1100 hPa
Please note that inhalation solutions intended for nebulization should also be at room temperature. If the
ambient temperature is too low (< 10° C) or too high (> 30° C), the nebulizer performance and the aerosol
spectrum of the SaHoMa®-II inhalation set for horses can no longer be guaranteed.
For tips on using the SaHoMa®-II inhalation set for hoses at cold temperatures, please read
chapter 3.6
“Performing an inhalation process at cold temperatures”.
7.4. Transport and storing conditions
In the following environmental conditions, the transport and storage of the components can take place without
causing damage to them or the components themselves representing a source of danger:
Ambient temperature
+ 10° C to +30° C; 20° C optionally
Relative air humidity
25 % to 75 %
Air pressure
450 hPa to 1100 hPa
The components must be stored and transported away from direct sunlight.
Do not store the components in damp rooms or transport them together with damp objects. In the event of
extreme fluctuations in the ambient temperature, the functionality of the components may be impaired by
condensation moisture.
Determination of the aerosol spectrum using a cascade impactor (NGI)*
The particle size is
3.92 µm MMAD
with a GSD of 3.17.
42.64 % of the particles are between 1.5 and 8 µm.
* The determination of the aerosol spectrum using a cascade impactor (NGI) was carried out by the Fraunhofer Institute for Toxicology and
Experimental Medicine ITEM in Hanover (Germany).
Note
These are an exemplary measurements of a representative sample.
MMD
: for Mass Mean Diameter [µm]
MMAD
: for Mass Median Aerodynamic Diameter [µm]
GSD
: Geometric Standard Deviation
Service
On our website you will find all kinds of information worth knowing about this and other topics related to inha
-
lation. All information is written by veterinarians in a simple and understandable way and kept in short form to
support you in the best possible way during the already nerve-racking time.