42

43

Attention

The use of 

self-mixed medications

 is always at the user‘s own risk, as both the mixing ratio and the  

ingredients are unknown and therefore proper functioning cannot be guaranteed.

The inhalation of 

essential oils or other plant extracts

 is generally not recommended with an electric nebuli-

zer, as these are too aggressive for nebulization and can lead to enormous irritation of the mucous membranes, 

up to and including severe allergic reactions and bronchospasms. In addition, they promote greatly premature 

wear of the SaHoMa®-II mesh nebulization unit.

Tests with various inhalation solutions have shown that chemical reactions and thus 

premature wear

 occur 

with subsequent preparations, so that their use is not recommended:

 

•  Essential oils

 

•  Preparations with the active ingredient acetylcysteine

 

•  Preparations with hypochlorous acid

Defects resulting from the use of these preparations do not justify any guarantee or warranty rights on your part.

Please note that although medications with different trade names may contain the same active ingredient,  

the dosage or composition may be different. Accordingly, the effect of the preparation on the components of the 

SaHoMa®-II inhalation set for horses is different, which in turn could influence the wear or the nebulization duration.

Warning

As a rule, nebulization of inhalation solutions suitable for inhalation with an electric nebulizer does not present 

any problems. It should always be noted that the SaHoMa®-II mesh nebulization unit is a consumable item that 

is subject to mechanical wear. The degree of wear depends on several factors, such as the inhalation solution, 

the inhalation frequency and quantity, and the care of the consumable. In the interest of effective inhala-

tion therapy, the SaHoMa®-II mesh nebulization unit must be replaced regularly, regardless of the inhalation  

solution used.

When using the above-mentioned inhalation solutions, please also adhere to the following recommendations 

with regard to the use and cleaning of the SaHoMa®-II mesh nebulization units:

 

•  Use one SaHoMa®-II mesh nebulization unit per inhalation solution and horse to avoid interactions 

or cross-contamination. For this purpose, label the SaHoMa®-II mesh nebulization unit accordingly to 

ensure traceability.

 • 

 Always use NaCl 0.9 % - not (sterile / distilled) water or brine - for mixing or diluting inhalation  

solutions to avoid the risk of clogging the SaHoMa®-II mesh nebulization unit with coarser salt crystals.

 

•   The  inhalation solution used should be mixed 1:1 with NaCl 0.9 % (e.g. 2 ml inhalation solution and  

2 ml NaCl 0.9 %), and at best even filled up to the maximum filling quantity of 17 ml (i.e. 2 ml inhala-

tion solution with 15 ml NaCl 0.9 %).

 

•   Mix  the inhalation solution in the medication and mixing cup shortly before use and not directly in the 

SaHoMa®-II mesh nebulization unit. Shake the medication and mixing cup well afterwards so that the 

inhalation solution combines with the NaCl 0.9 %. If you do not start nebulization after this step, shake 

the inhalation solution again immediately before filling and starting to prevent the preparations from 

separating again.

 

•   For nebulization, it is advisable to try it out in program 1 for the time being. If the SaHoMa®-II control 

unit switches off unintentionally in program 1 due to excessive bubble or foam formation, please start 

the SaHoMa®-II control unit in program 2 (time program). Set the time according to the inhalation  

volume used (guide value: approx. 1 ml per minute) and observe the inhalation process. If the  

SaHoMa®-II control unit does not switch off after nebulizing the inhalation solution, please switch it 

off manually at the end of the process. For further information, please refer to 

chapter 3 „Performing 

and ending the inhalation process“.

Cleaning the SaHoMa®-II mesh nebulization unit

The SaHoMa®-II mesh nebulization unit used should be cleaned with sterile or distilled water after each inhalation 

process. When using brine inhalation solutions or medications, a different cleaning procedure is necessary. For the 

exact cleaning procedure, please refer to 

chapter 4.2 „Cleaning the SaHoMa®-II mesh nebulization unit“.

6|

 Application instructions & Self-help

Application instructions & Self-help 

|6

Содержание SaHoMa-II HM-2000-1

Страница 1: ...set for horses Website Facebook Instagram Youtube Download center Manufacturer s information Manufacturer NEBU TEC International med Produkte Eike Kern GmbH Kreuzfeldring 17 63820 Elsenfeld Germany Ph...

Страница 2: ...ector 1 2 3 4 Notes 1 4 2 3 6 5 8 9 11 12 1 2 1 2 4 3 SaHoMa II housing HM 211 1 1 pc SaHoMa II protection cover HM 211 2 1 pc SaHoMa II filter basket HM 212 1 pc SaHoMa II filter membranes HM 214 6 p...

Страница 3: ...the following signs Note Information marked as Note should be given special attention Caution Information marked as Caution prevents damage to the product Warning Information marked with Warning is f...

Страница 4: ...e inhalation process 24 3 5 Putting down the SaHoMa II inhalation set for horses 25 3 6 Performing an inhalation process at cold temperatures 26 TABLE OF CONTENTS 4 Cleaning and storing the SaHoMa II...

Страница 5: ...veterinarian immediately Getting started with the SaHoMa II inhalation set for horses 2 Getting started with the SaHoMa II inhalation set for horses Read the instructions for use carefully before init...

Страница 6: ...ut a cut out and cut it completely individually After use put the SaHoMa II sealing ring in the outer compartment of the SaHoMa II carrying box for faster drying Getting started with the SaHoMa II inh...

Страница 7: ...t time it is recommended to adjust directly on the horse s head After initial adjustment there is no need to adjust again the next time the halter is put on Only minimal readjust ment may be required...

Страница 8: ...would result in too much aerosol escaping which may have a negative impact on the effectiveness of the inhalation therapy and thus the health of the horse 1 2 2 6 Using the SaHoMa II mesh nebulization...

Страница 9: ...lid fixes in different colors You can purchase this separately Art no 202 000 12 Getting started with the SaHoMa II inhalation set for horses 2 2 7 Filling the SaHoMa II mesh nebulization unit Cautio...

Страница 10: ...r draw it up completely in a syringe and fill it in from there Please note that the maximum filling quantity of the SaHoMa II mesh nebulization unit is 17 ml Now place the lid on the SaHoMa II mesh ne...

Страница 11: ...sockets of the SaHoMa II control unit 2 9 Sliding on the SaHoMa II protection cover Finally slide the SaHoMa II protection cover along the slide rails of the SaHoMa II housing over the SaHo Ma II mesh...

Страница 12: ...er the set time has elapsed An automatic switch off when the medication is used up does not occur in program 2 In chapter 3 2 Switching on the SaHoMa II control unit in program 1 and chapter 3 3 Switc...

Страница 13: ...wise the SaHoMa II control unit will display an error image For more information on error images please refer to chapter 6 3 Self help Troubleshooting Caution To activate the automatic switch off the...

Страница 14: ...the inhalation process by two light bars running synchronously from the center to the right and left After some time the SaHoMa II control unit switches off automatically The SaHoMa II control unit c...

Страница 15: ...inhalation You can also stop the inhalation process in Program 1 and Program 2 by manually switching off the SaHoMa II control unit To do this shortly press the button until the multifunction light o...

Страница 16: ...mperature to avoid rapid discharge and to ensure constant performance To avoid problems during nebulization we also recommend storing the SaHoMa II mesh nebuliza tion unit and the inhalation solution...

Страница 17: ...the SaHoMa II control unit and start it in program 1 by pressing the button Now wait until the filled sterile water has been completely nebulized Then switch off the SaHoMa II control unit 5 Remove th...

Страница 18: ...into the medication chamber of the SaHoMa II mesh nebulization unit and nebulize it as described in step 2 7 Remove the SaHoMa II mesh nebulization unit from the SaHoMa II control unit and allow the...

Страница 19: ...in the washing machine if it is very dirty Place the SaHoMa II halter strap in a padded laundry net and use the delicate cycle with a temperature of 30 C Use a mild detergent and avoid fabric softener...

Страница 20: ...n process It is recommended also for reasons of space to store in the SaHoMa II carrying box only those components that are regularly required for inhalation Additional consumables should be stored el...

Страница 21: ...be sufficient for a complete inhalation process In addition the battery capacity may also affect the quality of the inhalation process To charge the SaHoMa II control unit connect it to the SaHoMa II...

Страница 22: ...attery but not via the SaHoMa II power cord and the 230 VAC power supply socket When the lithium polymer battery is fully charged charging stops automatically to prevent the battery from overcharging...

Страница 23: ...sotonic solutions the mineral concentration is the same as in human cells in hypertonic solutions the mineral concentration is higher and in hypotonic solutions it is lower than in cells This causes i...

Страница 24: ...cleaning of the SaHoMa II mesh nebulization units Use one SaHoMa II mesh nebulization unit per inhalation solution and horse to avoid interactions or cross contamination For this purpose label the SaH...

Страница 25: ...t is already functional with the SaHoMa II mesh nebulization units changeover from February 2014 can have software updates installed on their SaHoMa II control unit and thus keep the SaHoMa II control...

Страница 26: ...ion of foaming medications according to the list of medications nebulization in program 2 under constant obser vation with manually switching off the SaHoMa II control unit Connect a SaHoMa II mesh ne...

Страница 27: ...he aerosol spectrum using a laser diffractometer HELOS VARIO KF The average particle size is 5 14 m MMD with a GSD of 1 91 48 28 of the particles are 5 m Technical data 7 1 Technical data of the SaHoM...

Страница 28: ...20 C optionally Relative air humidity 25 to 75 Air pressure 450 hPa to 1100 hPa The components must be stored and transported away from direct sunlight Do not store the components in damp rooms or tr...

Страница 29: ...g the SaHoMa II filter membrane just like a handkerchief depending on the degree of soiling and frequency of use The SaHoMa II filter membranes are not designed for cleaning If the end of the lifespan...

Страница 30: ...working hours The guarantee period starts from the date of purchase When purchasing our products from one of our distribu tors a receipt showing the date of purchase must be presented as proof of pur...

Страница 31: ...erefore cannot be removed from the old device 11 Options for returning old devices for consumers You can return or collect old devices free of charge within the framework of the possibilities for retu...

Страница 32: ...se needs how often and in what quantity The entry of the date gives you a history of all medications used You can use the last column for your own notes on the respective medication Date Medication Am...

Страница 33: ...Notes 60 Space for your notes...

Отзывы: