background image

46

47

6|

 FAQ & self help

FAQ & self help 

|6

Error

Possible Causes / Elimination

 

The inhalation times are

  

noticeably longer

 and there is 

no or only 

a small aerosol 

output.

Possible Causes:

•     Dirty or clogged membrane of the SaHoMa®-II mesh 

nebulization unit

•    Mechanically damaged membrane of the SaHoMa®-II mesh 

nebulization unit (e. g. in case of improper handling)

•   The end of the lifespan of the SaHoMa®-II mesh nebulization unit 

has been reached or exceeded

• The room or ambient temperature is below 10° C

Elimination:

•    Cleaning the SaHoMa®-II mesh nebulization unit 

(chapter 4.2)

•   Replacement of the SaHoMa®-II mesh nebulization unit

(chapter 9)

•   Please wait approx. 30 minutes until the system has adjusted to 

room temperature and only fill and nebulize inhalation solutions 

that are at room temperature.

•    Connect a SaHoMa®-II mesh nebulization unit filled with isotonic 

saline solution (NaCl 0.9 %), allow the device to nebulize for 

10 seconds and then switch it off again manually and switch it on 

again in the desired program

The SaHoMa®-II control unit  

switches off

 

unintentionally

  

during the inhalation process despite  

a remaining residual quantity.

 

Possible Causes:

•   Air bubbles on the membrane of the  SaHoMa®-II mesh 

nebulization unit

•  Use of a foaming medication suggesting a filling in form 

of foam residues

Elimination:

•  Light tapping of the SaHoMa®-II
•  mesh nebulization unit to loosen the air bubbles; 

if necessary restart the SaHoMa®-II control unit

•  Proper dilution of foaming medications according to the list 

of medications; nebulization in program 2 under constant obser-

vation with manually switching off the SaHoMa®-II control unit.

•  Connect a SaHoMa®-II mesh nebulization unit filled with isotonic

saline solution (NaCl 0.9 %), allow the device to nebulize for 10 

seconds and then switch it off again manually and switch it on 

again in the desired program

Error

Possible Causes / Elimination

Charging

 the

SaHoMa®-II control unit  

is 

not possible.

Possible Causes:

•  Defective SaHoMa®-II power cord or defective plug of the

 SaHoMa®-II power cord

•  No voltage at the socket used
•  Defective charging socket

Elimination:

•  Use of another socket
•  Check the plug connection at the SaHoMa®-II control unit
•  Otherwise send in for inspection/repair

The SaHoMa®-II control unit  

cannot be switched off 

and  

does not respond to pressing  

the button during operation.

Possible Causes:

•    The button on the SaHoMa®-II control unit has been pressed 

repeatedly within a short period of time

•     The  SaHoMa®-II control unit no longer responds to a button press 

due to an unusually high workload

Elimination:

•   Put the SaHoMa®-II control unit aside for approx. 10 minutes until 

it automatically switches itself off and performs a reset

Warning

During this time, the SaHoMa®-II control unit must not be 

connected for charging!

Service

In all other cases or in case of uncertainty, please contact our service staff who will be happy to help you:  

+49 (0) 6022 - 610 62 -0.

Please note that in the case of a repair, you will have to complete the repair form and send it to us with all the 

necessary components. Instructions for return can be found in the repair form.

Содержание SaHoMa-II HM-2000-1

Страница 1: ...set for horses Website Facebook Instagram Youtube Download center Manufacturer s information Manufacturer NEBU TEC International med Produkte Eike Kern GmbH Kreuzfeldring 17 63820 Elsenfeld Germany Ph...

Страница 2: ...ector 1 2 3 4 Notes 1 4 2 3 6 5 8 9 11 12 1 2 1 2 4 3 SaHoMa II housing HM 211 1 1 pc SaHoMa II protection cover HM 211 2 1 pc SaHoMa II filter basket HM 212 1 pc SaHoMa II filter membranes HM 214 6 p...

Страница 3: ...the following signs Note Information marked as Note should be given special attention Caution Information marked as Caution prevents damage to the product Warning Information marked with Warning is f...

Страница 4: ...e inhalation process 24 3 5 Putting down the SaHoMa II inhalation set for horses 25 3 6 Performing an inhalation process at cold temperatures 26 TABLE OF CONTENTS 4 Cleaning and storing the SaHoMa II...

Страница 5: ...veterinarian immediately Getting started with the SaHoMa II inhalation set for horses 2 Getting started with the SaHoMa II inhalation set for horses Read the instructions for use carefully before init...

Страница 6: ...ut a cut out and cut it completely individually After use put the SaHoMa II sealing ring in the outer compartment of the SaHoMa II carrying box for faster drying Getting started with the SaHoMa II inh...

Страница 7: ...t time it is recommended to adjust directly on the horse s head After initial adjustment there is no need to adjust again the next time the halter is put on Only minimal readjust ment may be required...

Страница 8: ...would result in too much aerosol escaping which may have a negative impact on the effectiveness of the inhalation therapy and thus the health of the horse 1 2 2 6 Using the SaHoMa II mesh nebulization...

Страница 9: ...lid fixes in different colors You can purchase this separately Art no 202 000 12 Getting started with the SaHoMa II inhalation set for horses 2 2 7 Filling the SaHoMa II mesh nebulization unit Cautio...

Страница 10: ...r draw it up completely in a syringe and fill it in from there Please note that the maximum filling quantity of the SaHoMa II mesh nebulization unit is 17 ml Now place the lid on the SaHoMa II mesh ne...

Страница 11: ...sockets of the SaHoMa II control unit 2 9 Sliding on the SaHoMa II protection cover Finally slide the SaHoMa II protection cover along the slide rails of the SaHoMa II housing over the SaHo Ma II mesh...

Страница 12: ...er the set time has elapsed An automatic switch off when the medication is used up does not occur in program 2 In chapter 3 2 Switching on the SaHoMa II control unit in program 1 and chapter 3 3 Switc...

Страница 13: ...wise the SaHoMa II control unit will display an error image For more information on error images please refer to chapter 6 3 Self help Troubleshooting Caution To activate the automatic switch off the...

Страница 14: ...the inhalation process by two light bars running synchronously from the center to the right and left After some time the SaHoMa II control unit switches off automatically The SaHoMa II control unit c...

Страница 15: ...inhalation You can also stop the inhalation process in Program 1 and Program 2 by manually switching off the SaHoMa II control unit To do this shortly press the button until the multifunction light o...

Страница 16: ...mperature to avoid rapid discharge and to ensure constant performance To avoid problems during nebulization we also recommend storing the SaHoMa II mesh nebuliza tion unit and the inhalation solution...

Страница 17: ...the SaHoMa II control unit and start it in program 1 by pressing the button Now wait until the filled sterile water has been completely nebulized Then switch off the SaHoMa II control unit 5 Remove th...

Страница 18: ...into the medication chamber of the SaHoMa II mesh nebulization unit and nebulize it as described in step 2 7 Remove the SaHoMa II mesh nebulization unit from the SaHoMa II control unit and allow the...

Страница 19: ...in the washing machine if it is very dirty Place the SaHoMa II halter strap in a padded laundry net and use the delicate cycle with a temperature of 30 C Use a mild detergent and avoid fabric softener...

Страница 20: ...n process It is recommended also for reasons of space to store in the SaHoMa II carrying box only those components that are regularly required for inhalation Additional consumables should be stored el...

Страница 21: ...be sufficient for a complete inhalation process In addition the battery capacity may also affect the quality of the inhalation process To charge the SaHoMa II control unit connect it to the SaHoMa II...

Страница 22: ...attery but not via the SaHoMa II power cord and the 230 VAC power supply socket When the lithium polymer battery is fully charged charging stops automatically to prevent the battery from overcharging...

Страница 23: ...sotonic solutions the mineral concentration is the same as in human cells in hypertonic solutions the mineral concentration is higher and in hypotonic solutions it is lower than in cells This causes i...

Страница 24: ...cleaning of the SaHoMa II mesh nebulization units Use one SaHoMa II mesh nebulization unit per inhalation solution and horse to avoid interactions or cross contamination For this purpose label the SaH...

Страница 25: ...t is already functional with the SaHoMa II mesh nebulization units changeover from February 2014 can have software updates installed on their SaHoMa II control unit and thus keep the SaHoMa II control...

Страница 26: ...ion of foaming medications according to the list of medications nebulization in program 2 under constant obser vation with manually switching off the SaHoMa II control unit Connect a SaHoMa II mesh ne...

Страница 27: ...he aerosol spectrum using a laser diffractometer HELOS VARIO KF The average particle size is 5 14 m MMD with a GSD of 1 91 48 28 of the particles are 5 m Technical data 7 1 Technical data of the SaHoM...

Страница 28: ...20 C optionally Relative air humidity 25 to 75 Air pressure 450 hPa to 1100 hPa The components must be stored and transported away from direct sunlight Do not store the components in damp rooms or tr...

Страница 29: ...g the SaHoMa II filter membrane just like a handkerchief depending on the degree of soiling and frequency of use The SaHoMa II filter membranes are not designed for cleaning If the end of the lifespan...

Страница 30: ...working hours The guarantee period starts from the date of purchase When purchasing our products from one of our distribu tors a receipt showing the date of purchase must be presented as proof of pur...

Страница 31: ...erefore cannot be removed from the old device 11 Options for returning old devices for consumers You can return or collect old devices free of charge within the framework of the possibilities for retu...

Страница 32: ...se needs how often and in what quantity The entry of the date gives you a history of all medications used You can use the last column for your own notes on the respective medication Date Medication Am...

Страница 33: ...Notes 60 Space for your notes...

Отзывы: