Trouble-
shooting
27
GERMAN
Lower Aerosol
output
(remaining quantity is
too high)
1.
Used or damaged
medication cup.
1. Replace
medication
cup.
2. Contact fluid level in
contact fluid reservoir
too high/low.
2.
Fill contact fluid
reservoir with 45
ml
distilled water (use
measuring cup with
markings).
3. Contact fluid reservoir
not properly cleaned.
3. Clean device according
to instructions.
4. Several
medication
cups used.
4.
Only use one
medication cup.
Device doesn’t
generate
any aerosol
1. Several
medication
cups used.
1.
Only use one
medication cup.
2.
Used or damaged
medication cups used.
2.
Use new medication
cups.
3.
Device is not
connected to power
supply.
3.
Connect device to
power supply.
4. No contact fluid filled in
contact fluid reservoir.
4.
Fill contact fluid
reservoir up to the
correct level.
5. No fluid (medication
solution) in the
medication cup.
5. Fill medication cup.
Inhalation or
exhalation
impeded
1.
Filter membrane is
clogged or saturated.
1. Replace
filter
membrane.
2. Nebuliser top piece is
not properly fastened.
2. Check if the nebulizer
top piece is properly
fastened.