Initial operation
12
GERMAN
5.10 Use of OPTINEB
®
-ir in the ventilation system with climate-control
filter
Can only be used with ventilators with the medication nebulization option.
If you are using climate-control filters or an active humidification for your
ventilation, please proceed as specified below.
Insert the filter housing [
A17
] including an inserted filter membrane [
A9.1
] into the
top piece [
A4
] on the side.
Unscrew the white Luer/Lock plug [
A7
] and place it on the parking space
intended for it [
A5
]. Connect the control line for mechanical nebulization [
A22
]
with the adapter of your ventilator intended for this purpose and with the
Luer/Lock/connection [
A6
] of the
OPTINEB
®
-ir
ultrasonic nebulizer.
Now connect the
OPTINEB
®
-ir
[
A1
] ultrasonic nebulizer by adaption with the
hose system for the aerosol inlet close to the patient [
A19
] with the ventilator
hose system (T-piece between climate-control filter and tube).
(sterile hose system incl. control line, angle piece, 60cm creased tube and T-
piece) [
A19
].
5.11 Use of OPTINEB
®
-ir in the ventilation system with coaxial hoses
Can be used with ventilators with the medication nebulization option.
If you
are using coaxial hoses for your ventilation, please proceed as specified below.
Insert the filter housing [
A9
] including an inserted filter membrane [
A9.1
] into the
top piece [
A4
] on the side.
Filter
Expiration
Inspiration
Patient
Mechanical nebulisation A22 control line