Power supply
15
GERMAN
6.3.1
The battery is charged through the power supply [
A17
]. For this, connect
the plug of the power supply to the battery [
A16
]).
6.3.2
The charging time for the battery is approx. 8–10 hours.
6.3.3
Charging should never exceed 12 hours.
6.3.4
Battery operation is not possible while the battery is charging.
6.3.5
The battery must be disconnected from the power supply of the
OPTINEB
®
-ir
after charging.
6.3.6
Connect the charged battery only for the duration of the inhalation with the
OPTINEB
®
-ir
.
6.3.7
Please pull out the battery plug from the device after ending inhalation.
6.3.8
Only after the display [
A12
] of the
OPTINEB
®
-ir
shows the letter
combination (LB) may the battery be charged again.
When the battery charge is 100%, operation of the ultrasonic nebulizer
OPTINEB
®
-ir
of approx. 40 min. is possible.
NOTE during battery operation
Do not use the device while bathing.
The device is to be set up so that it cannot fall into water.
Do not immerse the device into water or any other fluid.
Do not use the device if it has fallen into water.
Immediately pull out the battery plug.
CAUTION
In order to prevent damage to the ultrasonic nebulizer and ensure adherence to
EMC guidelines, only the original power supply [
A17
] may be used.
NOTE
Drop off defective battery cells for disposal at battery disposal sites or return
them to NEBU-TEC GmbH.