Initial operation
13
GERMAN
Unscrew the white Luer/Lock plug [
A7
] place it on the parking space intended for
it [
A5
]. Connect the control line for mechanical nebulization [
A22
] to the adapter
of your ventilator intended for this purpose and with the Luer/Lock/connection of
the
OPTINEB
®
-ir
ultrasonic nebulizer.
A climate-control filter can be used at the inspiratory end of the coaxial hose in
the direction of the patient.
Now connect the
OPTINEB
®
-ir
[
A1
] ultrasonic nebulizer by adaption with the
hose system for the aerosol inlet close to the patient [
A19
] with the ventilator
hose system (T-piece between climate-control filter and tube).
(sterile hose system incl. control line, angle piece, 60cm creased tube and T-
piece) [
A19
].
IMPORTANT
Use only sterile medication cups [
A10
] for use in the ventilator.
When using the
OPTINEB
®
-ir
in the ventilation system, use only sterile
accessories.
Aerosols only come into contact with the autoclaved parts and not with the actual
device.
In order to protect your circle part from aerosol residue and your ventilator
against damage, an end expiratory mechanical filter must be used
(manufacturing company: Hudson, Pall, B+P, Tyco).
Compatibility checks are available for the most widely-used ventilators.
Filter
Expiration
Inspiration
Patient
Mechanical nebulisation A22 control line