5.0 Initial operation of your OPTINEB
®
-ir ultrasonic nebulizer for
ventilation
Preparing the ultrasonic nebuliser.
5.01
Ensure first that the
OPTINEB
®
-ir
[
A1
] ultrasonic nebulizer is
not connected to a power source.
Remove any power supply by pulling out
the connection cables from the socket.
(Reverse side of
OPTINEB
®
-ir
).
5.02
Take the
OPTINEB
®
-ir
ultrasonic nebulizer and fill it with 45 ml
of distilled
or demineralized
water.
ATTENTION
The use of other contact fluids (e.g. tap water, sterile water or saline solution) is
strictly prohibited since this can lead to significant impairment of the performance
of the device and even to complete malfunction.
5.03
Insert a sterile medication cup [
A10
] with the
tip facing downwards into the ultrasonic
nebulizer. Bear in mind that the medication
cup must be submerged in the contact fluid.
5.04
Now begin to assemble the autoclaved and
therefore sterile plastic pieces. Begin with the top piece
[
A4
]. Check whether the baffle plate is firmly seated in
the housing, the sealing ring is in its guide and the white
Luer/Lock plug is securely locked.
5.05
Now place the top piece [
A4
] on the
ultrasonic nebulizer and turn it once on its
own axis until a light click can be heard. Do
not force when turning it.