8.2 Description of the ultrasonic nebulizer OPTINEB
®
-ir 2.4 MHz
program
The following statements apply for programs P1 to P5:
The main function of the keys
“Start/Stop”
and
“On/Off”
are identical.
In program 6 (P6), the
“Start/Stop”
key is neutralized.
The selection of the program is described in Point “8.1 Programming mode
only for authorized persons”
8.2.1 Features of the first program (P1)
Program 1 was developed for the nebulization of special medications.
Non-adjustable nebulization time: max. 12 minutes. Time indication on the
display runs from “0” going forward until the pre-set time is reached. The aerosol
is intermittently generated (no continuous aerosol production). After expiry of the
pre-set time, the program is ended.
8.2.2 Features of the second program (P2)
Program 2 was developed for the nebulization of special medications.
Non-adjustable nebulization time: max. 12 minutes.
Time indication on the display runs from “0” going forward until the pre-set time is
reached. The aerosol is intermittently generated (no continuous aerosol
production). After expiry of the pre-set time, the program is ended. The user is
not able to change the program parameters.
8.2.3 Features of the third program (P3)
No fixed nebulization time. The device is volume-controlled (remaining quantity
recognition) and produces aerosol until the medication has been nebulized. The
OPTINEB
®
-ir
ultrasonic nebulizer switches off automatically after reaching a
remaining quantity of approx. 0.5 ml. The inhalation time may differ in length and
results from the set ventilation parameters, the respiratory rate and the depth of
respiration