Maintenance
25
GERMAN
In-house hygiene instructions with regard to the replacement interval must be
complied with as a matter of priority.
NOTE
The medication cups are disposable containers and must
be replaced for hygiene and technical reasons daily (see
Section
10.1). Non-adherence to the prescribed
replacement intervals can lead to deformation of the
medication cups [
A10
]. These micro-deformations of the
medication cup [
A10
] can lower the output of the
ultrasonic nebulizer considerably.
12.0 Servicing
The
OPTINEB
®
-ir ultrasonic nebulizer
should be serviced every 2 years.
Servicing may only be performed by NEBU-TEC GmbH or a specifically
authorized or qualified NEBU-TEC dealer.
WARNING
Do not open the housing. Non-observance results in the warranty being voided.