11 / 45
General Information
| 11
4
General Information
4.1
Device Requirements
4.1.1 System Safety Requirements
External equipment connected to the signal input/output or other connectors of this product for use in a pa-
tient environment must comply with the requirements of ISO and ANSI/AAMI ES/EN/IEC 60601-1 safety stand-
ards. A person who connects such equipment to this product has by definition formed a system, and is re-
sponsible for compliance of that system to the same ISO and ANSI/AAMI ES/EN/IEC 60601-1 safety standards.
NDS recommends that VIMA installation be conducted by qualified personnel.
4.1.2 Power Supply
This product complies to the listed safety standards only when used with the supplied medical grade power
supply:
Model
Radiance® Ultra 32" 4K ZeroWire Duo
Power Supply
BridgePower BPM150S24F06
AC Input
100 - 240 V, 50 to 60 Hz
DC Output
24 Vdc, 6.25 A, 150 W
4.1.3 Power Cord
Use the supplied hospital grade power cord with the correct plug for your power source.
• The power cord is the only recognized disconnect device for this product. To power off the product, dis-
connect the power cord from the AC mains.
• The product and other medical equipment should be positioned so that the power cord and connection to
AC mains is readily accessible.
• If an extension cord or power strip is needed for connection of this product to AC mains, confirm that the
power cord plug can be securely connected to the cord or power strip. The low voltage side of the power
supply can only be extended to up to 50 ft.
• This product should be powered from a center tapped circuit when used in the US at voltages over 120 V.
4.1.4 Grounding
This product is energized from an external electrical power source for 2017/745 Medical Device Regulation of
European Union Class I equipment. It is the responsibility of the installer to test the product's earth ground to
verify that it complies with the hospital, local and national impedance requirements.
A ground post is located on the back of the product to use for grounding the chassis of the unit. Any such
grounding must be installed in accordance with applicable electrical codes.