27 / 45
Set up and Installation
| 27
Elevation
The primary and secondary display should be positioned at least 5 feet (1.5 m) above the floor, be at the same
height, and, preferably, with the primary and secondary display fronts parallel and facing each other.
Horizontal distance between displays
Radiance® Ultra 32" 4K ZeroWire Duo
operates correctly with a horizontal distance between the displays of
up to 30 feet (9.1 m). However, in the case of most OR environments, best results are achieved with a hori-
zontal distance of up to 8 feet (2.4 m). To avoid loss of image or flickering, minimize line of sight obstruction
between primary and secondary display.
Min. 5 ft (1.5 m)
Max. 8 ft (2.4 m)
Min. 5 ft (1.5 m)
(1)
(2)
(5)
(4)
(3)
Positioning of displays
(1)
Radiance® Ultra 32" 4K ZeroWire Duo
,
primary display
(4)
Surgical table
(2)
Radiance® Ultra 32" 4K ZeroWire Duo
, sec-
ondary display
(5)
Roll-stand
(3)
Endoscopy cart