User and Service Manual
EMU40EX™
19
Electromagnetic Interference (EMI) Warnings
Only use the EMU40EX system in conjunction with approved devices and accessories. Use of cables other than
those specified or sold by the manufacturer on the equipment, may result in increased emissions or decreased
immunity of the equipment and may cause the system to be non-compliant with the requirements of IEC
60601-1-2:2007.
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
To ensure the validity of signals, do not operate the device near any sources of electromagnetic interference.
The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent
or stacked use is necessary, the equipment or system should be observed to verify normal operation in the
configuration in which it will be used
Transportation Warnings
Make sure that any platform, table, cart, or other surface used during the operation, transport, or temporary or
permanent storage of the system and its components is adequate, sturdy, and safe. Natus is not responsible for
any injury or damage that may result from inadequate, poorly constructed, or unapproved transports, carts, or
operating surfaces. Natus is not responsible for any injury or damage that may result from improper cable storage
during transport
Do not tilt the trolley/cart more than 10° incline as this will compromise the stability of the trolley/cart.
TIPPING HAZARD: During transport, the user should guide the cart using both hands, ensuring the wheel base is
aligned so that a single caster leads in the direction of motion. Failure to lead the cart with one wheel could result
in a tipping hazard when ascending or descending steps or thresholds.
WARNING: Only use Natus approved equipment on the trolley/cart. Non-approved equipment may compromise
the function and safety of the system.
Do not leave the amplifier attached to the computer when transporting the unit.
Cleaning Warnings
Do not immerse patient cables in water, solvents, or cleaning solutions (the patient cable connectors are not
waterproof). Do not sterilize patient cables by irradiation, steam or ethylene oxide.
Do NOT soak or immerse the amplifier in any liquid.
Be careful not to allow any fluid to seep into the internal electronic components of the system.