A pulse oximeter should be considered an early warning device. As a trend towards patient deoxygenation is
indicated, blood samples should be analyzed by a laboratory CO-oximeter to completely understand the
patient’s condition.
The EMU40EX Amplifier system does NOT include SpO2 or Pulse Rate alarms.
Wireless Option Warnings
NOTE
: The EMU40EX contains FCC ID: SQGBT700 which complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to
operate the equipment.
NOTE
: The EMU40EX contains a Class 1 Bluetooth module which has passed the Bluetooth Qualification/Certification process as
specified within the Bluetooth Specifications.
NOTE
: There are no devices known to cause disturbance to the performance of the wireless radios in this device. However,
precautions should be made to keep the EEG device a safe distance from any other wireless products. Refer to Table 4 in the
Declaration of Compliance for IEC 60601-1-2 section. Should interference with the wireless radios in this device occur, relocate
the devices causing the issue, or contact Natus Technical Support at 1-800-303-0306 or
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in this manual and any other Accompanying Documents.
The wireless feature of this amplifier/headbox presents a passive visualization of the EEG signal. This feature is
not a substitute for other means of patient monitoring and supervision.
Like other wireless devices, this feature is limited in operating range. The range will also be reduced
substantially if there is any physical interference between the Base Unit and the Breakout Box. In either of
these situations, the wireless data will not be received by the acquisition computer.
The EMU40EX contains a Class 1 Bluetooth module which has passed the Bluetooth Qualification/Certification
process as specified within the Bluetooth Specifications.
There are no devices known to cause disturbance to the performance of the wireless radios in this device.
However, precautions should be made to keep the EEG device a safe distance from any other wireless
products. Refer to Table 4 in the Declaration of Compliance for IEC 60601-1-2 section.
Should interference with the wireless radios in this device occur, relocate the devices causing the issue, or
contact Natus Technical Support at 1-800-303-0306 or [email protected].