User and Service Manual
EMU40EX™
55
Discrete Saturation Transform (DST™) plot
The DST plot has two peaks: the peak corresponding to the higher saturation is selected as the SpO
2
value. This entire sequence
is repeated once every two seconds on the most recent four seconds of raw data. The Masimo SET
®
pulse oximeter SpO
2
therefore corresponds to a running average of arterial hemoglobin saturation that is updated every two seconds.
Pulse Oximeter Accessories
The Pulse Oximeter Accessories, PC Series Cable and LNCS series Oximetry sensors are sold separate from the EMU40EX
System.
•
PC Series Cable
: Connects the pulse oximeter module to the oximetry sensors. Transfers LED drive power to the
oximetry sensors from the pulse oximeter. The pulse oximeter receives the detector signal from the oximetry sensor
through the PC series cable.
•
LNCS
®
Series Oximetry Sensors
: Low Noise Cabled Sensors (LNCS) and integrated cables are designed to provide
optimal pulse oximetry performance. Offering additional protection from liquids and improved moisture resistance,
these sensors perform accurately with motion, low perfusion, and intense ambient light. LNCS
®
sensors can be used
with all Masimo SET
®
pulse oximeters.
Available Oximeter Sensors
The inbuilt Pulse Oximeter from Massimo is compatible with following accessories to be purchased directly from Masimo.
Please contact your local Masimo reseller directly to purchase the following approved sensors:
Model Number
Description
1006 (PC-12)
MASIMO PATIENT CABLE 12' - CONNECTS TO ALL MASIMO PULSE OXIMETER SENSORS
2013
Masimo LNCS Reusable 14ft Patient Cable
2653
Masimo LNCS DBI Reusable Adult Soft with 3ft Cable
1859
Masimo LNCS Disposable Adult Sensor with 18" Cable