EMU40EX™
User and Service Manual
14
The sale, distribution, or use of this device is restricted to, by, or on order of a physician.
Printers and/or peripherals used with Natus devices should be marked and approved for use in their appropriate
medical environment in accordance with the regulations of the country in which they are used.
Do not attempt to connect any third-party devices to the Patient Breakout Box except those approved by Natus.
WARNING: No modification of this equipment is allowed.
WARNING: Misaligned connector pins can cause damage to system components at power-on.
WARNING: A damaged cable can cause a short in the electrical circuit. Prevent damage to the connectors by
aligning connector pins before you connect the cable.
Changes or modifications not expressly approved by the manufacturer could void the user's authority to operate
the equipment.
Perform the recommended maintenance. Refer to the Maintenance, Cleaning, & Disposal section for details.
The Natus EMU40EX with the Natus Base is NOT designed to work with defibrillators. The system could be
damaged when used with this device.
EMU40EX Amplifiers are compatible with NeuroWorks / SleepWorks v9.0 and newer.
Operation of this equipment with input signals in excess of a range of +/-10mV may cause incorrect results.
REPETITIVE STRESS INJURY HAZARD: Sustained use of this product without ergonomic consideration may result in
repetitive stress injury.
Do NOT position Medical Electrical equipment in such a way as to make it difficult to operate the disconnection
device.
Do not dispose of this medical product as unsorted municipal waste. At the end of an EMU40EX's life, please
contact Natus at 1-800-303-0306 in order to arrange return shipping for your EMU40EX amplifier.
Electrical Warnings and Cautions
Natus systems are intended for connection to a properly grounded electrical outlet only. Ensure that the
EMU40EX Amplifier System is solely connected to a three-wire, grounded, hospital-grade receptacle.
The use of accessories, transducers and/or cables other than those specified, with the exception of those sold by
the manufacturer as replacement parts for internal components, may result in increased emissions or decreased
immunity of the equipment or system and may cause the system to be non-compliant with the requirements of
IEC 60601-1-2:2007.