User and Service Manual
EMU40EX™
15
Conductive parts of electrodes and their connectors are not to contact other conductive parts and earth.
Do not place MULTIPLE PORTABLE SOCKET-OUTLETS (MPSOs) on the floor.
Do not connect additional MPSOs or extension cords to the EMU40EX Amplifier System.
WARNING: ELECTRICAL SHOCK HAZARD: Do NOT turn on the system power until all cables have been connected,
verified and visually inspected for any damage. Failure to inspect the cables may result in electrocution.
Verification of electrical safety should be performed routinely.
ELECTRICAL SHOCK HAZARD: Do NOT service the system. Refer servicing to qualified personnel only. Do NOT use
repaired components without proper testing.
Do not use the MPSO with the EMU40EX Amplifier System for supplying power to any equipment that is not part
of the system.
Do NOT connect electrical equipment which has not been supplied as a part of the system to the MPSO.
ELECTRICAL SHOCK HAZARD: Do NOT connect electrode inputs to earth ground. The Patient Breakout Box contains
warning symbols reminding you that the connections are intended for isolated patient connections only.
Connecting to an earth ground might result in electrocution.
Do NOT connect non-medical equipment which has been supplied as part of the system directly to the wall outlet
when the system is supplied, via MPSO, with a separating transformer.
To avoid the possible hazards caused by the summation of leakage currents when all the parts of the system are
interconnected, no equipment other than devices connected to the EMU40EX Amplifier system may be powered
by the isolation transformer.
The current rating of the isolation transformer must be sufficient to operate all of the devices powered by it. Refer
to the current ratings of the isolation transformer and current rating for each individual device connected
When replacing the fuse for the EMU40EX, it must be replaced by a fuse with the same type and rating as the
original fuse. Replacement fuses should be purchased from Natus directly.
The Patient Breakout Box accepts only touch-proof style electrode inputs. Do NOT attempt to use any other style
of patient electrode input.
For the use of an ELECTROENCEPHALOGRAPH without protective means to prevent burning of the PATIENT,
information on the location of the ELECTRODES is provided.