5.4.3
The TEOAE test
If
Child Mode
is activated, an animated car is displayed on the screen during testing.
(See
.)
The progress bar under the animated picture is green if the test conditions are good
and it is red if there is too much noise. Noise can come from the environment or the
patient.
If
Child Mode
is not enabled, test data is shown on the display as the test progresses.
•
If needed, press the
Stop
button to stop the test. The test will be saved as an
incomplete test.
Test details
Shows which ear is being tested.
A.
Valid peaks
B.
Higher frequency emissions
C.
Lower frequency emissions
The graphic display shows the peaks
reaching the test criteria. The graph is a
statistical graph.
Valid peak
Shows that the intensity of a peak is valid and counted by the peak counter.
Otometrics - MADSEN Alpha OAE+
35
5 Testing with Alpha OAE+