Voltage dips, short inter-
ruptions and voltage vari-
ations on power supply input
lines IEC 61000-4-11
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase: at 0°
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase: at 0°
Mains power quality should be that of a typical com-
mercial or hospital environment. If the user of the
MADSEN Alpha OAE+ requires continued operation dur-
ing power mains interruptions, it is recommended that
the MADSEN Alpha OAE+ be powered from an unin-
terruptible power supply or a battery.
Voltage interruptions on
power supply input lines
IEC 61000-4-11
0 % U
T
; 250/300 cycles
0 % U
T
; 250/300 cycles
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
30 A/m
No relevant ports that could be
affected
Power frequency magnetic fields should be at levels char-
acteristic of a typical location in a typical commercial or
hospital environment.
U
T
is the AC mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems within Professional Healthcare use environment
MADSEN Alpha OAE+ is intended for use in the electromagnetic environment specified below. The user of MADSEN Alpha OAE+ should ensure that it is used in
such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz to 80 MHz
6 V rms
IISM Bands
3 V rms
150 kHz to 80 MHz
6 V rms
IISM Bands
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
Proximity fields from RF wire-
less communications
IEC 61000-4-3
27 V/m
386 MHz
28 V/m
450 MHz,
9 V/m
710 MHz, 745 MHz, 780 MHz
28 V/m
810 MHz, 870 MHz, 930 MHz,
28 V/m
1720 MHz, 1845 MHz, 1970 MHz
28 V/m
2450 MHz,
9 V/m
5240 MHz, 5500 MHz, 5785 MHz
27 V/m
386 MHz
28 V/m
450 MHz,
9 V/m
710 MHz, 745 MHz, 780 MHz
28 V/m
810 MHz, 870 MHz, 930 MHz,
28 V/m
1720 MHz, 1845 MHz, 1970 MHz
28 V/m
2450 MHz,
9 V/m
5240 MHz, 5500 MHz, 5785 MHz
Separation distance between any electronic parts of
MADSEN Alpha OAE+ and any RF wireless com-
munication equipment must be more than 30 cm (11.8
inches).
Note
: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
80
Otometrics - MADSEN Alpha OAE+
App. 3 Technical Specifications