Protocol 1 and Protocol 2
6 kHz
5 kHz
4 kHz
3.5 kHz
3 kHz
2.5 kHz
Minimum DP amplitude equals -5 dB.
A
Pass/Clear Response
result requires that 4 of the 6 frequencies receive a
PASS
result.
Protocol 1 stops automatically if 4 frequencies reach a
Pass/Clear Response
result or 3 frequencies receive a
Refer/No
Clear Response
result.
Protocol 2 tests all frequencies.
Protocol 3 and 4
5 kHz
4 kHz
3 kHz
2 kHz
Minimum DP amplitude equals -5 dB.
A
Pass/Clear Response
result requires that 3 of the 4 frequencies receive a
Pass/Clear Response
result.
Protocol 3 stops automatically if 3 frequencies reach a
Pass/Clear Response
result or 2 frequencies receive a
Refer/No
Clear Response
result.
Protocol 4 tests all frequencies.
Protocol 5 and 6
6 kHz
5 kHz
4 kHz
3 kHz
2 kHz
1.5 kHz
Minimum DP amplitude equals -5 dB.
Protocol 5: a
Pass/Clear Response
result requires that 4 of the 6 frequencies receive a
Pass/Clear Response
result. The
test stops automatically if 4 frequencies reach a
Pass/Clear Response
result or 3 frequencies receive a
Refer/No Clear
Response
result.
Protocol 6: no overall
Pass/Clear Response
or
Refer/No Clear Response
result is generated. All frequencies are tested.
28
Otometrics - MADSEN Alpha OAE+
5 Testing with Alpha OAE+