Standards
Safety:
EN 60601-1:2006+A1:2013
ANSI/AAMI ES60601-1:2005 + A1:2012
CAN/CSA-C22.2 NO. 60601-1:14
Internally Powered, Type BF, IPX0
IEC 60601-2-40:2016 and EN 60601-2-40:1998
EMC:
IEC 60601-1-2:2007 and EN 60601-1-2:2007
IEC 60601-1-2:2014 and EN 60601-1-2:2015
Otoacoustic emissions:
IEC 60645-6:2009, Type 2 and EN 60645-6:2010, Type 2
App. 3.4
Power supply and battery
Device
Supply voltage:
Nom. 3.70 V,
Max. 4.20 V,
Min. 3.20 V (measured with device load)
Maximum battery power consumption:
1.5 W while measuring
Estimated battery life:
8 hours of continuous use (based on a typical use scenario. Actual use can influ-
ence the battery life time.)
Battery level indicator:
5-step battery level indicator
Charge time in Alpha OAE+ docking sta-
tion:
80% charged: 4½ hours
Battery
Caution
•
Use only rechargeable battery supplied by Natus Medical Denmark ApS, Part no. 8-73-02400. Use of any
other battery may present a risk of fire or explosion.
Battery type:
Rechargeable Li-ion 3.7 V/1800 mAh (6.7 Wh), fully charged
App. 3.5
1077 docking station
PC interface
Interface type:
USB 2.0, Full-speed
USB Power:
Uses <100 mA of current from the USB interface