Test details
Peak counter
The test must register a total of at least 8 valid peaks in alternating directions
(counted both above and below the median line) in order to lead to a
Pass/Clear Response
.
Noise
Current noise level. Make sure the
Noise
level is as low as possible. The
Noise
bar turns red when the noise level is over 41 dB SPL. A high
Noise
level may
indicate a high level of environmental noise or a noisy patient.
Artifact
Shows the average artifact level in percent. The artifact rate (emanating from
noise) should be below 20%, if possible.
If artifacts are present, this may indicate a high level of environmental noise or a
noisy patient. In these cases, check the test conditions and make sure that the
probe fits.
Stability
Shows the average probe stability level in percent. The stimulus stability (eman-
ating from probe stability) should be above 80%, if possible.
Progress
Shows the time elapsed for the current test frequency. If a
Pass/Clear Response
for the current frequency is not achieved the test frequency will time out. Max.
time depends on environmental noise conditions. If test conditions are optimal
but there is no response measured, then the frequency will time out before the
progress bar reaches the end.
5.4.4
TEOAE test results
In Child Mode, when a test is completed, the Reward screen is displayed, regardless of the test result. This screen shows
an animated bunch of balloons floating up to the sky. The outcome of the test is then displayed on the
Test result
screen.
If Child Mode is not activated, the
Test result
screen is displayed immediately after the test is completed.
•
Press the
OK
button to return to the
Test Menu
.
36
Otometrics - MADSEN Alpha OAE+
5 Testing with Alpha OAE+