4
Preparing for testing
4.1
Preparing Alpha OAE+
•
Follow the instructions in
Getting Started with Alpha OAE+
•
See
on how to proceed with the specific tests.
4.2
Preparing the equipment
Every day before you start testing on patients, you should test the probe to make sure that it functions correctly. See
Caution
•
Insert only disinfected probe tips in the test cavity. See
Cleaning and disinfecting the built-in test cavity
4.2.1
Preparing the probe for testing
Inspect the probe for deterioration (color changes, surface changes) of the reusable probe parts before every usage. If
deterioration occurs, contact your distributor.
4.2.2
Connecting the probe
1. Place Alpha OAE+ face down.
2. Align the ridge on the probe plug with the notch in the grey-bordered socket on the top end of Alpha OAE+.
3. Gently insert the probe plug into the probe socket. The plug does not require force to be inserted properly.
Disconnecting the probe
Caution
•
Do not pull the plug by the cable when you disconnect the probe. Instead, pull the sleeve of the grey con-
nector.
22
Otometrics - MADSEN Alpha OAE+
4 Preparing for testing