ICS Aircal
ICS Aircal system safety
Verify that the patient is positioned properly, in the supine position with head flexed 30
°
.
5.2
Testing procedure
Perform each of the four irrigations as follows:
Verify that a new speculum has been placed on the delivery head.
Instruct the patient to keep eyes open (VNG) or closed (ENG).
Close goggles for VNG. Have patient close their eyes for ENG.
Start collecting response and start irrigation.
Start tasking patient 20-30 seconds into irrigation.
When duration time is reached and the ICS Aircal device beeps, remove the delivery head from the patient's ear and
continue tasking.
Right after the peak of the response, allow the patient to fixate.
Collect eye movements until the response subsides.
Dispose of the speculum and carefully place the delivery head in its cradle.
If the ICS Aircal device is connected to an ICS VNG/ENG system, subsequent depressions of the flow time activation
switch will center the tracing for VNG/ENG or start the video recording for VNG. (If installed, the foot switch will also
serve to start and center tracings or start the video recording.)
6
ICS Aircal system safety
This guide contains information, cautions, and warnings that must be followed to ensure the safe performance of the
ICS Aircal system.
When the ICS Aircal device is used in conjunction with a VNG/ENG Systems, make sure that all information, cautions,
and warnings in the User Manual for the test device are followed. For safety specifics concerning VNG/ENG Systems,
see the specific manuals.
Any serious incident that has occurred in relation to the device should be reported to Natus Manufacturing Limited
and the competent authority of the Member State in which the user and/or patient is established.
Note
•
Local government rules and regulations, if applicable, should be followed at all times.