ICS Aircal
ICS Aircal Front Panel
3.5
Airflow Rate
3.5.1
Recommended and default Airflow Rate
The default airflow rate for caloric irrigation is 8 liters per minute.
3.5.2
Setting a new Airflow Rate
Turn the knob to the right of the airflow
button to adjust the flow rate. Turn the knob counter clockwise to
increase the airflow and clockwise to decrease the airflow. The range for the airflow rate is 4.0 to 10 liters per minute.
The airflow rate is set based on the knob position.
The flow rate selection (actual only) will be indicated on the display panel. If the ICS Aircal device is in operation mode,
the digital display shows the actual flow rate. If the ICS Aircal device is in standby mode, the display will be off.
3.6
Beep volume
A beep will sound when the timer reaches 0 to alert the tester to stop irrigation. A 2000 Hz beep is presented.
3.6.1
Default Beep setting
The default beep is set to level 2.
3.6.2
Setting a new Beep volume
To adjust the beep volume, press and hold the beep
button until you hear a beep that you want as the
default. There are four levels for the beep volume. To change the default value of the beep volume, see
3.7
Changing the Stored Default Values
Change all settings to the desired setting. Decide the start temperature for the ICS Aircal. Press and hold either the
warm
or cool
temperature button for 3 seconds to set all default settings (warm and cool
temperatures, startup temperature, duration timer, and beep volume). A 1-second beep will sound to confirm that
the default settings have been stored. Remember, the default startup temperature of the ICS Aircal device should
match the temperature of the first caloric protocol in your list (e.g. if you start with Left Cool, default the ICS Aircal
device to start with Cool).
3.8
Water level display
The water level in the water reservoir is displayed. The
symbol indicates that the water reservoir is full and a blue
bar will display. The
symbol indicates that the water reservoir is half full and a blue bar will display. Between half
full and empty the bar will change from blue to orange, which indicates that the water reservoir needs to be refilled.
The
symbol indicates that the water reservoir is empty.